FDA: Don't use CardioGen-82
On July 15, FDA alerted the medical imaging community about the potential for inadvertent, increased radiation exposure in patients who underwent or will be undergoing cardiac PET scans with Rb-82 chloride injection from CardioGen-82.
The FDA had received reports of two patients who received more radiation than expected from CardioGen-82. The agency is suggesting that the risk of harm from this exposure is minimal, though exposure to any excessive radiation is undesirable. The estimated amount of excess radiation the two patients received is similar to that other patients may receive with cumulative exposure to certain other types of heart scans, according to FDA.
Based on further investigation, the agency has determined that the current CardioGen-82 manufacturing procedures are not sufficient to ensure reliable performance of the generator used to produce the Rb-82 chloride injection. Reliable generator performance is essential to help prevent strontium breakthrough from CardioGen-82 and to prevent patients from being exposed to excess radiation. The FDA also is currently investigating the sufficiency of the testing procedures used to detect strontium breakthrough at the clinical sites that use CardioGen-82.
The agency is recommending that healthcare professionals use alternatives to the CardioGen-82 generator when planning nuclear medicine cardiac scans. Patients who have any questions or concerns should talk to their healthcare professional.
FDA continues to work with the Nuclear Regulatory Commission and the CardioGen-82 manufacturer to determine the root cause for the increased radiation exposure detected in the two patients. The extent to which any additional patients may have received inadvertent radiation exposure is also under investigation. FDA plans to notify the public with updates.
For more information please visit: CardioGen82.