Magnetic eyelashes aren’t so pretty on MRI scans

Magnetic eyelashes were the top beauty-related Google search term last year, but the faux lashes can have adverse effects if worn during an MRI scan.

In fact, such eyelashes can reduce the quality of an image and even put the patient in danger, wrote authors Einat Slonimsky, MD, and Alexander Mamourian, MD, of Penn State Health Milton S. Hershey Medical Center in Hershey, Pennsylvania.

“Although friction and adhesion may differ from patient to patient, depending on the width and character of the native eyelashes of an individual,” the pair wrote, “we strongly recommend inserting a line about magnetic eyelashes on the MRI safety questionnaire and adding stops in the screening system to prevent the entry of anyone with these lashes, including staff, into the MRI scanner room.”

For the study, published July 24 in the American Journal of Roentgenology, the team bought two sets of randomly selected magnetic eyelashes and attached them to nylon strings located within a water-based phantom. The phantom was submerged in distilled water before going through a variety of MRI scan types. These results were compared to images gathered using an aneurysm clip submerged in the water.

Overall, the eyelashes crated a “substantial artifact,” much larger than that created by the aneurysm clips, the authors noted. The artifact was 7 × 6 cm, completely blocking the entire phantom.

“Although we tested only one kind of magnetic eyelashes, it is reasonable to assume that all such lashes will behave in this manner, producing either somewhat less or somewhat more magnetic field distortion, and that all will be attracted to the static magnet,” the researchers wrote.

""

Matt joined Chicago’s TriMed team in 2018 covering all areas of health imaging after two years reporting on the hospital field. He holds a bachelor’s in English from UIC, and enjoys a good cup of coffee and an interesting documentary.

Around the web

The nuclear imaging isotope shortage of molybdenum-99 may be over now that the sidelined reactor is restarting. ASNC's president says PET and new SPECT technologies helped cardiac imaging labs better weather the storm.

CMS has more than doubled the CCTA payment rate from $175 to $357.13. The move, expected to have a significant impact on the utilization of cardiac CT, received immediate praise from imaging specialists.

The newly cleared offering, AutoChamber, was designed with opportunistic screening in mind. It can evaluate many different kinds of CT images, including those originally gathered to screen patients for lung cancer. 

Trimed Popup
Trimed Popup