Boston Sci launches Class I recall of IVUS catheters

Boston Scientific has commenced a voluntary recall of nearly 30,000 imaging catheters that were distributed in the U.S. and U.S. territories, after multiple confirmed cases of catheter tip detachments in the iCross coronary imaging catheters. The devices, used in intravascular ultrasound (IVUS) in patients who are candidates for transluminal coronary interventional procedures, are part of an FDA Class I recall.

The corrective action, announced May 27, is being taken due to eight confirmed cases of catheter tip detachments caused by the embrittlement of catheter material. The Natick, Mass.-based company confirmed a rate of 0.027 percent of catheter tip detachments in the U.S. and Puerto Rico from April 1, 2010, to May 10, 2011.

The catheters are intended to operate with Boston Scientific’s ClearView Ultra, Galaxy, Galaxy2 or iLab IVUS imaging consoles.

Potential health risks associated with this type of failure include vessel wall injury, thrombotic events, retained foreign body, foreign body embolization, myocardial infarction and death, according to the company. There are additional risks to health associated with retrieval attempts. The majority of confirmed brittle tip detachments have been successfully retrieved, according to the press release.

Boston Scientific said it is notifying affected hospitals through detailed recall notification letters, including instructions on how to return the recalled product, which also are located on the company’s website.

Around the web

Richard Heller III, MD, RSNA board member and senior VP of policy at Radiology Partners, offers an overview of policies in Congress that are directly impacting imaging.
 

The two companies aim to improve patient access to high-quality MRI scans by combining their artificial intelligence capabilities.

Positron, a New York-based nuclear imaging company, will now provide Upbeat Cardiology Solutions with advanced PET/CT systems and services.